BL983313-1 Conductivity Mini Controller (0 - 1999 ¦ÌS/cm)
Previous | 5 / 16 | Next |
The BL983313 is a compact, panel mounted, process controller for measuring electrolytic conductivity (EC) of a process stream that is within the 0 to 1999 μS/cm range. Users may choose from automatic or manual dosing modes. When in automatic mode the dry contact relay is activated when a reading is above the set point. The relay can be used to activate a dosing pump or a solenoid that controls a valve.
Multicolor LED Indicator
LED changes color based on status.
Fire Retardant Casing
For additional safety the controllers outer case is made of fire resistant plastic.
Splash Resistant Cover
A clear plastic cover protects the face panel from dirt and liquids.
This compact, panel mounted, process controller is for measuring electrolytic conductivity (EC) of a process stream. Users may choose from automatic or manual dosing modes. When in automatic mode the dry contact relay is activated when a reading is above the set point.
HI983313's relay can be used to activate a dosing pump or a solenoid that controls a valve.
HI983313 is ideal for source water or rinse water applications.
Adjustable Dry Contact Dosing Relay
The BL983313 features a dosing relay that is activated when the reading is above a user programmable set point.
Programmable Overdose Protection
For enhanced safety the mini controller can be programmed to deactivate the dosing relay if the set point is not reached within a specified time interval. The overdosing timer is programmable from 5 to 30 minutes or disabled.
Relay Control Override
With the flick of a switch the mini controller's relay can be disabled (Off), placed in control (Auto), or be activated for manual operation (On) which is useful for priming a dosing pump.
Multicolor LED Indicator
Multicolor LED indicator allows an operator to quickly check the status of the controller.
- Green = Meter in measurement mode and reading is below the set point.
- Orange/Yellow = Reading is above the set point and the relay is activated.
- Blinking Red = Indicates an alarm condition such as when the maximum dosing time has been exceeded.
Fuse Protected Dosing Contacts
The relay dosing contact is rated for up to a 2A load and is fuse protected.
Labeled Termination Connections
Quick-connect terminal blocks are clearly labeled for easy connection to power, conductivity probe and relay that can be used to operate a dosing pump, valve, audible alarm, or light.
- Model BL983313-2 is available with 4-20 mA Galvanic Isolated Output with external dosing disable contact
- Large Clear LCD
- Fire Retardant Casing
- Splash-Resistant Cover
1/2'' NPT Male Threaded Amperometric Probe for Mounting In-line*
The BL983313 mini controller uses the HI7634 conductivity probe that is available with different cable lengths.
*HI7634 is not included and must be ordered separately
Front View
Installation cut-out dimensions are provided as well as outside panel dimensions.
Side View
Adjustable location brackets allow the controller to slide easily and securely into place.
SKU | BL983313-web |
---|---|
Product Name | Conductivity Mini Controller (0 - 1999 μS/cm) - BL983313 |
Quote Required | Yes |
Range | 0 to 1999 µS/cm |
Resolution | 1 µS/cm |
Accuracy | ±2% f.s. at 25 °C (77 °F) |
Temperature Compensation | automatic from 5 to 50°C (41 to 122°F) with β =2%/ºC |
Calibration | manual, with calibration trimmer |
Output | galvanic isolated 4-20 mA output; accuracy ±0.2 mA; |
Output | 500 Ω maximum load (BL983313-2 only) |
Dosing adjustable setpoint | covers measure range |
Dosing relay | relay closes when reading > setpoint |
Dosing dosing contact | maximum 2A (internal fuse protection),250 VAC or 30 VDC |
Dosing overtime | Dosing relay is disabled if setpoint is not reached within the set time interval. Timer adjustable between approx. 5 to 30 minutes, or disabled by jumper. |
Dosing external disable input | Normally Open |
Power Supply | models “-0” - 12 VDC adapter (included) |
Power Supply | models “-1” 115/230 VAC; 50/60Hz |
Power Supply | models “-2” |
Power Supply | input |
Dimensions | 83 x 53 x 99 mm (3.3 x 2.1 x 3.9’’) |
Weight | 12 VDC models, 200 g (7.1 oz); 115/230 VAC models 300 g (10.6 oz) |